Show simple item record

dc.contributor.authorBanerjee, K.
dc.contributor.authorGhosh, T. K.
dc.contributor.authorKundu, S.
dc.contributor.authorRana, T. K.
dc.contributor.authorBhattacharya, C.
dc.contributor.authorMeenaa, J. K.
dc.contributor.authorMukherjee, G.
dc.contributor.authorMali, P.
dc.contributor.authorGupta, Dhruba
dc.contributor.authorMukhopadhyay, S.
dc.contributor.authorPandit, D.
dc.contributor.authorBanerjee, S. R.
dc.contributor.authorBhattacharya, Sabita
dc.contributor.authorBandyopadhyay, T.
dc.contributor.authorChatterjee, S.
dc.date.accessioned2012-11-29T10:45:33Z
dc.date.available2012-11-29T10:45:33Z
dc.date.issued2009-09-21
dc.identifier.citationBanerjee K, Ghosh T K, Kundu S, Rana T K, Bhattacharya C, Meena J K, Mukherjee G, Mali P, Gupta D, Mukhopadhyay S, Pandit D, Banerjee S R, Bhattacharya S, Bandyopadhyay T, . Chatterjee S (2009) Variation of neutron detection characteristics with dimension of BC501 A neutron detector, Nucl. lnstr. Meth. A, 608, 440.en_US
dc.identifier.issn0168-9002
dc.identifier.uri1. Full Text Link ->
dc.identifier.urihttp://www.sciencedirect.com/science/article/pii/S0168900209014788#en_US
dc.identifier.uri=================================================en_US
dc.identifier.uri2. Scopus : Citation Link ->en_US
dc.identifier.urihttp://www.scopus.com/record/display.url?eid=2-s2.0-69449088688&origin=resultslist&sort=plf-f&src=s&st1=Variation+of+neutron+detection+characteristics&st2=Gupta%2cD&sid=vt4Gq2k9s1V1prHImmw6gj2%3a660&sot=b&sdt=b&sl=93&s=%28TITLE-ABS-KEY-AUTH%28Variation+of+neutron+detection+characteristics%29+AND+AUTHOR-NAME%28Gupta%2cD%29%29&relpos=0&relpos=0&searchTerm=%28TITLE-ABS-KEY-AUTH%28Variation%20of%20neutron%20detection%20characteristics%29%20AND%20AUTHOR-NAME%28Gupta,D%29%29en_US
dc.identifier.uriFOR ACCESS / DOWNLOAD PROBLEM -- PLEASE CONTACT LIBRARIAN, BOSE INSTITUTE, akc@bic.boseinst.ernet.inen_US
dc.descriptionDOI: 10.1016/j.nima.2009.07.034en_US
dc.description.abstractA systematic study of the variation of neutron detection characteristics (efficiency, pulse shape discrimination) and the intrinsic time resolution with the active volume of the detector has been carried out with liquid scintillator (BC501A)-based neutron detectors of various dimensions designed and fabricated for this purpose. Energy-dependent neutron detection efficiency has been measured using associated particle technique and its dependence on detector dimension has been studied. The measured efficiencies have been compared with those obtained from GEANT4 simulations.en_US
dc.language.isoenen_US
dc.publisherELSEVIER SCIENCE BVen_US
dc.subjectLiquid scintillator BC501Aen_US
dc.subjectEfficiency measurementen_US
dc.subjectSimulationen_US
dc.titleVariation of neutron detection characteristics with dimension of BC501A neutron detectoren_US
dc.title.alternativeNUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENTen_US
dc.typeArticleen_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record